The biggest updated is that we now our pivoting to our second highest Pugh scoring device - the veritical mnp syringe. It will be 1 ml, lowering the volume capabilities. This prevent the issue of forcing mnps to the centerline, which was difficult given the strength of feasible ems and the width of the syringe. The mount to hold the pump 90* is being designed and constructed by Zane and Bert. One Em will still be pulsed at the junction, for mixing at the junction.
This week the volume verification tests are being performed. The infusate motor is working to about +- 5% accuracy, so the accuracy spec will need to be increased. One to five percent error is normal for the market, especially for lower end pumps. This was accepted by our client.
There is a working user interface screen.
All spec changes were approved by client.